Please use this identifier to cite or link to this item: https://hdl.handle.net/11147/7608
Title: Determination of Arsenic by Hydride Generation—laser-Induced Breakdown Spectroscopy: Characterization of Interelement Interferences
Authors: Bölek, Deniz
Ünal Yeşiller, Semira
Yalçın, Şerife
Keywords: Aqueous samples
Arsenic
Hydride generation
Spectroscopy
Publisher: Taylor and Francis Ltd.
Source: Bölek, D., Ünal Yeşiller, S., and Yalçın, Ş. (2018). Determination of arsenic by hydride generation—laser-induced breakdown spectroscopy: Characterization of interelement interferences. Analytical Letters, 51(10), 1605-1621. doi:10.1080/00032719.2017.1384927
Abstract: In this study, interelement interferences were evaluated for the determination of arsenic in aqueous samples through laser-induced breakdown spectroscopy (LIBS) hyphenated with a hydride generation sample introduction system. Optimum instrumental and chemical parameters were selected and variation in LIBS signal intensity was recorded for As solution in the presence of comparable concentrations of interfering elements. No significant change in the signal intensity of As(I) 228.8 nm line was observed in the presence of alkali/alkali earth metals; however, the presence of hydride-forming elements has shown a noticeable decrease in the line emission strength of arsenic. The least variation in arsenic signal was observed in the presence of Ge, the most volatile of all. However, the signal has decreased to a greater extent in the presence of Sn, Sb, and Pb. The presence of interfering elements on electron temperature and electron number density of arsenic plasma has also been studied. Plasma temperatures calculated using both As and Ar emission lines in the Boltzmann equation were similar, being around 5000 K. The McWhirter criterion for stationary and homogenous plasmas was utilized for the establishment of the local thermodynamic equilibrium under the plasma conditions studied. Applicability of the technique for multielemental analysis of water samples was tested through spiking experiments. Arsenic signal showed 26% decrease in the multielemental mixture solution. LIBS is among a few atomic spectroscopic techniques that facilitate rapid and simultaneous multielemental analysis without extensive sample preparation steps. However, the analytical performance of the technique still requires more serious efforts to compete with other conventional techniques for routine analysis of environmental samples.
URI: https://doi.org/10.1080/00032719.2017.1384927
https://hdl.handle.net/11147/7608
ISSN: 0003-2719
Appears in Collections:Chemistry / Kimya
Scopus İndeksli Yayınlar Koleksiyonu / Scopus Indexed Publications Collection
WoS İndeksli Yayınlar Koleksiyonu / WoS Indexed Publications Collection

Files in This Item:
File Description SizeFormat 
Binder1.pdfMakale (Article)1.19 MBAdobe PDFThumbnail
View/Open
Show full item record



CORE Recommender

SCOPUSTM   
Citations

3
checked on Dec 20, 2024

WEB OF SCIENCETM
Citations

2
checked on Dec 21, 2024

Page view(s)

662
checked on Dec 16, 2024

Download(s)

404
checked on Dec 16, 2024

Google ScholarTM

Check




Altmetric


Items in GCRIS Repository are protected by copyright, with all rights reserved, unless otherwise indicated.