Title: | Size-based microparticle seperation using negative magnetophoresis | Authors: | Solmaz Özçelik, Özge Öksüz, Cemre Tekin, Hüseyin Cumhur |
Keywords: | Microfluidic filtration Negative magnetophoresis Size-based separation |
Publisher: | Chemical and Biological Microsystems Society | Abstract: | We present a new size-based microparticle separation device using negative magnetophoresis. Microparticles spiked in the paramagnetic medium were filtered with respect to their sizes in a microfluidic channel placed between two magnets. Negative magnetophoresis allows large microparticles to be captured before the magnets, while small microparticles pass through the magnets under a constant flow. With this method, we reached 84.2% capturing efficiency of large microparticles (44 µm diameter) and capturing purity of 80.3% in the presence of small microparticles (17 µm diameter) at 3 µL/min flow rate. The capturing purity could further improve up to 99% by increasing the flow rate. | URI: | https://hdl.handle.net/11147/12489 |
Appears in Collections: | Bioengineering / Biyomühendislik Scopus İndeksli Yayınlar Koleksiyonu / Scopus Indexed Publications Collection |