Please use this identifier to cite or link to this item: https://hdl.handle.net/11147/12317
Title: Effect of silicon nitride coating thickness on silicon wafer substrates for signal enhancement in laser-induced breakdown spectroscopic analysis of liquids
Authors: Kaplan, Dilara
Yalçın, Şerife Hanım
Keywords: Dried droplet analysis
Thermal conductivity
Silicon nitride
Laser-induced breakdown spectroscopy
Publisher: Elsevier
Abstract: It has been shown by previous studies of our group that the use of nitride-coated silicon wafer surfaces as a sample loading substrate in dried-droplet LIBS analysis provided enhancement in plasma emission signal and better detection limits compared to uncoated or oxide-coated silicon wafer surfaces. To further investigate the effect of coating thickness for enhanced sensitivity in dried nano-droplet analysis of liquids, silicon-wafer substrates of different nitride coating thicknesses; 75, 300, 450, and 1000 nm, were comparatively studied. With 75 nm silicon nitride coating, the thin-film effect due to the anti-reflective behavior of the silicon nitride film is observed, and plasma emission signal is enhanced up to three times compared to 300 nm coated substrates. With coating thicknesses of 450 nm and 1000 nm, on the other hand, thermophysical and mechanical properties of the silicon nitride material, like thermal conductivity and hardness, become more dominant factors, leading to higher emission signals for all the elements studied. With 1000 nm coating thickness, enhancement factors of 4.8, 6.4, and 3.7 were obtained for the elements of Pb, Cu, and Cr, respectively. Optimization of the experimental LIBS parameters was conducted, calibration curves were constructed, and analytical figures of merits were determined. Sub-picogram amounts absolute detection limits; 0.7 pg Pb, 0.6 pg Cr, and 0.4 pg Cu, in 500 nanoliter droplets were obtained from the slopes of the calibration curves. The nitride-coated substrates' analytical performance was tested using certified reference solutions, standard water, and real water samples. The materials and the methodology developed can be used for waste-water monitoring of environmental samples by LIBS.
Description: The authors acknowledge research grant 119F068 from TUBITAK
URI: https://doi.org/10.1016/j.sab.2022.106472
https://hdl.handle.net/11147/12317
Appears in Collections:Chemistry / Kimya
Scopus İndeksli Yayınlar Koleksiyonu / Scopus Indexed Publications Collection
WoS İndeksli Yayınlar Koleksiyonu / WoS Indexed Publications Collection

Files in This Item:
File Description SizeFormat 
1-s2.0-S0584854722001161-main.pdf
  Until 2024-05-01
Article (Makale)4.41 MBAdobe PDFView/Open    Request a copy
Show full item record



CORE Recommender

SCOPUSTM   
Citations

5
checked on Apr 5, 2024

WEB OF SCIENCETM
Citations

5
checked on Mar 30, 2024

Page view(s)

12,762
checked on Apr 15, 2024

Google ScholarTM

Check




Altmetric


Items in GCRIS Repository are protected by copyright, with all rights reserved, unless otherwise indicated.