dc.date.issued | 2006-04 | |
dc.identifier.citation | Türkan, U., Öztürk, O., and Eroğlu, A. E. (2006). Metal ion release from TiN coated CoCrMo orthopedic implant material. Surface and Coatings Technology, 200(16-17), 5020-5027. doi:10.1016/j.surfcoat.2005.05.005 | en_US |
dc.identifier.issn | 0257-8972 | |
dc.identifier.issn | 0257-8972 | - |
dc.identifier.uri | http://doi.org/10.1016/j.surfcoat.2005.05.005 | |
dc.identifier.uri | http://hdl.handle.net/11147/2218 | |
dc.description.abstract | In this study, a medical grade CoCrMo alloy was coated with TiN by means of physical vapor deposition (PVD) technique at 550 °C for 6 h. The TiN layer microstructure and thickness were studied by X-ray diffraction (XRD) and cross-sectional scanning electron microscopy (SEM). The adhesive strength of the TiN coatings on the CoCrMo substrate was studied by a commercially available Scanning Scratch Tester. Static immersion test was conducted to investigate the effectiveness of TiN coating in preventing the dissolution of metal ions into the simulated body fluid (SBF) from the substrate by atomic absorption spectrometry (AAS) and inductively coupled plasma optical emission spectrometry (ICP-OES). The XRD results showed that the PVD coated TiN films exhibited (111) preferred orientation, while the SEM analysis indicated quite uniform and highly dense TiN coated layer (about 3 μm thick) with a columnar growth mode reaching from substrate to coating surface. The scratch test results showed that the adhesive strength between the TiN film and the CoCrMo substrate was adequate. The AAS and ICP-OES results showed that the presence of the TiN coating prevented the release of cobalt and chromium metal ions from the substrate CoCrMo alloy whereas cobalt was preferentially dissolved from the as-polished material. Calcium phosphate precipitation was observed on the surface of the as-polished material, indicating a degree of bioactivity of the as-polished surface which is absent in the TiN coated substrate alloy. | en_US |
dc.description.sponsorship | IYTE 2003/14 | en_US |
dc.language.iso | en | en_US |
dc.publisher | Elsevier Ltd. | en_US |
dc.relation.ispartof | Surface and Coatings Technology | en_US |
dc.rights | info:eu-repo/semantics/openAccess | en_US |
dc.subject | Cobalt alloys | en_US |
dc.subject | Atomic absorption spectrometry | en_US |
dc.subject | Cobalt-chromium alloy | en_US |
dc.subject | Immersion test | en_US |
dc.subject | Scanning electron microscopy | en_US |
dc.title | Metal ion release from TiN coated CoCrMo orthopedic implant material | en_US |
dc.type | Article | en_US |
dc.authorid | TR4824 | en_US |
dc.authorid | TR8641 | en_US |
dc.institutionauthor | Türkan, Uğur | - |
dc.institutionauthor | Öztürk, Orhan | - |
dc.institutionauthor | Eroğlu, Ahmet Emin | - |
dc.department | İzmir Institute of Technology. Physics | en_US |
dc.department | İzmir Institute of Technology. Chemistry | en_US |
dc.identifier.volume | 200 | en_US |
dc.identifier.issue | 16-17 | en_US |
dc.identifier.startpage | 5020 | en_US |
dc.identifier.endpage | 5027 | en_US |
dc.identifier.wos | WOS:000236590300048 | en_US |
dc.identifier.scopus | 2-s2.0-33645012459 | en_US |
dc.relation.publicationcategory | Makale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanı | en_US |
dc.identifier.doi | 10.1016/j.surfcoat.2005.05.005 | - |
dc.relation.doi | 10.1016/j.surfcoat.2005.05.005 | en_US |
dc.coverage.doi | 10.1016/j.surfcoat.2005.05.005 | en_US |
dc.identifier.wosquality | Q1 | - |
dc.identifier.scopusquality | Q1 | - |
item.fulltext | With Fulltext | - |
item.grantfulltext | open | - |
item.languageiso639-1 | en | - |
item.openairecristype | http://purl.org/coar/resource_type/c_18cf | - |
item.cerifentitytype | Publications | - |
item.openairetype | Article | - |
crisitem.author.dept | 04.05. Department of Pyhsics | - |
crisitem.author.dept | 04.01. Department of Chemistry | - |
Appears in Collections: | Physics / Fizik Scopus İndeksli Yayınlar Koleksiyonu / Scopus Indexed Publications Collection WoS İndeksli Yayınlar Koleksiyonu / WoS Indexed Publications Collection |
CORE Recommender
SCOPUSTM
Citations
71
checked on Nov 15, 2024
WEB OF SCIENCETM
Citations
59
checked on Oct 26, 2024
Page view(s)
292
checked on Nov 18, 2024
Download(s)
596
checked on Nov 18, 2024
Google ScholarTM
Check
Altmetric
Items in GCRIS Repository are protected by copyright, with all rights reserved, unless otherwise indicated.